微量氧分析儀是一種常用的分析儀器。它是世界上的氧氣測量方法之一。它由高活性氧電極和鉛電極組成,浸在KOH溶液中。
The micro oxygen analyzer is a common analytical instrument. It is one of the most advanced oxygen measurement methods in the world. It is composed of high active oxygen electrode and lead electrode, immersed in KOH solution.
微量氧分析儀廣泛用于測定上、下塔的氮純度、精氬塔的微量氧含量等工藝控制參數。同時,還可以對輸氣管道、液氮、儲罐氬中微量氧含量等質量控制參數進行測量。根據儀器的不同測量原理,綜合考慮生產控制的質量要求、經濟成本、維護工作量和維護成本等因素,選擇微氧分析儀。
The micro oxygen analyzer is widely used to determine the nitrogen purity of the upper and lower towers, the micro oxygen content of the refined argon tower and other process control parameters. At the same time, it can also measure the quality control parameters such as the content of micro oxygen in gas pipeline, liquid nitrogen and storage tank argon. According to the different measuring principles of the instrument, the micro oxygen analyzer is selected considering the quality requirements, economic cost, maintenance workload and maintenance cost of production control.
微量氧分析儀在使用過程中難免會出現各種情況下的故障,下面就來介紹一下:
It is inevitable that the micro oxygen analyzer will fail in various situations during use. Here is an introduction:
(1)儀器故障表明計量槽溫度偏低。檢查量筒熱電偶的連接,檢查多通道轉接板的安全,檢查線路板的連接。
(1) The failure of the instrument indicates that the temperature of the metering tank is low. Check the connection of the measuring cylinder thermocouple, the safety of the multi-channel adapter plate, and the connection of the circuit board.
(2)儀器故障表示測量槽的高電位或低電位。檢查測量槽的電路連接,檢查電路板之間的連接,檢查氧化鋯傳感器。
(2) Instrument fault indicates high or low potential of measuring tank. Check the circuit connection of the measuring slot, check the connection between the circuit boards, and check the zirconia sensor.
(3)傳感器對樣品氣體含氧量變化響應緩慢,測量值較低。在2005年1月,一個氧化鋯微量氧分析儀被發現有一個非常低的測量和緩慢的傳感器響應。經過對儀器采樣系統的檢查、零點范圍的校準和傳感器的檢查,zui終確認傳感器存在故障。更換傳感器后,儀器恢復正常測量。
(3) The sensor responds slowly to the change of oxygen content of sample gas, and the measured value is low. In January 2005, a zirconia micro oxygen analyzer was found to have a very low measurement and slow sensor response. After the inspection of the instrument sampling system, the calibration of the zero range and the inspection of the sensor, Zui finally confirmed that the sensor was faulty. After replacing the sensor, the instrument returns to normal measurement.
(4)儀器工作環境中的氧含量應為空氣中的氧含量,以避免其他因素對工作環境氧含量的影響。
(4) The oxygen content in the working environment of the instrument shall be the oxygen content in the air to avoid the influence of other factors on the oxygen content in the working environment.
(5)儀器投入使用時,所有采樣管均應充入潔凈空氣或氮氣進行清洗;在儀器的進口處應安裝過濾器,以防止在空分裝置啟動時,采樣線上的雜質損壞傳感器。
(5) When the instrument is put into use, all sampling pipes shall be filled with clean air or nitrogen for cleaning; a filter shall be installed at the inlet of the instrument to prevent impurities on the sampling line from damaging the sensor when the air separation unit is started.
(6)保證儀表的排放管路有平滑的大氣壓,以避免壓力過大和排氣管堵塞,影響儀表的測量精度。
(6) Ensure that the discharge pipe of the instrument has smooth atmospheric pressure, so as to avoid excessive pressure and exhaust pipe blockage, which will affect the measurement accuracy of the instrument.
(7)被測樣品氣體中的可燃氣體對氧化鋯氧分析儀的測量結果有一定的影響,但可燃氣體含量(如甲烷、一氧化碳、二氧化碳等)非常低,對微量氧測量的干擾很小。
(7) The combustible gas in the sample gas has a certain influence on the measurement results of zirconia oxygen analyzer, but the combustible gas content (such as methane, carbon monoxide, carbon dioxide, etc.) is very low, which has little interference on the measurement of micro oxygen.